Caller ID Signal Levels (Part 2: Standards)

Caller ID Signal Levels (Part 3: Standards)

Introduction:

The Caller ID levels stated in various standards vary significantly in how they are defined. Along with the use of different units, these standards may specify unique termination conditions. The following table provides a summary of signal level definitions for three common Caller ID standards. These being Telecordia (Bellcore) SR-3004, Telecommunication Industry Association (TIA) TIA-716 & TIA-777, and European Telecommunications Standards Institute (ETSI) ETS 300 778-1 & ETS 300 778-2.

Summary of Level Definitions:

 

CPE State

Standard

On-Hook

Off-Hook

Telecordia

(Bellcore)

Document: SR-3004
Units: dBm (referenced to 600 ohms)
Source Impedance: 900 ohms
Load Impedance: 600 ohms

Document: SR-3004
Units: dBm (referenced into 600 ohms)
Source Impedance: 900 ohms
Load Impedance: 600 ohms

 

Setup: Levels are measured across tip and ring with the line terminated by a 600 ohm resistor. When performing the tests, replace the 600 ohm resistor by the CPE.

Setup: Levels are measured across tip and ring with the line terminated by a 600 ohm resistor. When performing the tests, replace the 600 ohm resistor by the CPE.

 

Example: Sending a FSK signal level of -13.5 dBm requires a CO voltage source level of -5.54 dBm (0.409 Vrms). The level across tip and ring during testing is also -5.54 dBm (0.409 Vrms).

Example: Sending a FSK signal level of -13.5 dBm requires a CO voltage source level of -5.54 dBm (0.409 Vrms). The level across tip and ring during testing will depend on the CPE’s off-hook impedance. Since this is normally close to 600 ohms, the level will be approximately -13.5 dBm (0.163 Vrms).

TIA

Document: TIA/EIA-716
Units: mVrms
Source Impedance: 900 ohms
Load Impedance: open circuit

Document: TIA/EIA-777
Units: mVrms
Source Impedance: 900 ohms
Load Impedance: 600 ohms

 

Setup: Levels are measured across tip and ring with the line open circuit. When performing the tests, connect the CPE to the telephone line.

Setup: Levels are measured across tip and ring with the line terminated by a 600 ohm resistor. When performing the tests, replace the 600 ohm resistor by the CPE.

 

Example: Sending a FSK signal level of 120 mVrms requires a CO voltage source level of 120 mVrms. The level across tip and ring during testing is also 120 mVrms.

Example: Sending a FSK signal level of 120 mVrms requires a CO voltage source level of 300 mVrms. The level across tip and ring during testing will depend on the CPE’s off-hook impedance. Since this is normally close to 600 ohms, the level will be approximately 120 mVrms.

ETSI

Document: ETS 300 778-1
Units: dBV
Source Impedance: not specified
Load Impedance: the CPE’s AC termination can be either:
a) greater than 8 kohms
b) impedance of 820 ohms in series with a parallel combination of 360 ohms and 180 nF.

Document: ETS 300 778-2
Units: dBV
Source Impedance: not specified
Load Impedance: CPE’s AC termination requirements depend on additional national standards

 

Setup: Levels are measured across tip and ring with the CPE connected in the on-hook state (off-line).

Setup: Levels are measured across tip and ring with the CPE connected in the off-hook state (on-line).

 

Example: Sending a FSK signal level of -20 dBV requires adjusting the CO voltage source until the level across tip and ring is -20 dBV. The actual CO sending level depends on the CPE’s AC termination and the source impedance of the test setup.

Example: Sending a FSK signal level of -20 dBV requires adjusting the CO voltage source until the level across tip and ring is -20 dBV. The actual CO sending level depends on the CPE’s AC termination and the source impedance of the test setup.