|
CPE State |
Standard |
On-Hook |
Off-Hook |
Telecordia
(Bellcore) |
Document: SR-3004
Units: dBm (referenced to 600 ohms)
Source Impedance: 900 ohms
Load Impedance: 600 ohms
|
Document: SR-3004
Units: dBm (referenced into 600 ohms)
Source Impedance: 900 ohms
Load Impedance: 600 ohms
|
|
Setup: Levels are measured across tip and ring with the line
terminated by a 600 ohm resistor. When performing the tests,
replace the 600 ohm resistor by the CPE.
|
Setup: Levels are measured across tip and ring with the line
terminated by a 600 ohm resistor. When performing the tests,
replace the 600 ohm resistor by the CPE.
|
|
Example: Sending a FSK signal level of -13.5 dBm requires a
CO voltage source level of -5.54 dBm (0.409 Vrms). The level
across tip and ring during testing is also -5.54 dBm (0.409 Vrms). |
Example: Sending a FSK signal level of -13.5 dBm requires a
CO voltage source level of -5.54 dBm (0.409 Vrms). The level
across tip and ring during testing will depend on the CPE’s
off-hook impedance. Since this is normally close to 600 ohms,
the level will be approximately -13.5 dBm (0.163 Vrms).
|
TIA |
Document: TIA/EIA-716
Units: mVrms
Source Impedance: 900 ohms
Load Impedance: open circuit
|
Document: TIA/EIA-777
Units: mVrms
Source Impedance: 900 ohms
Load Impedance: 600 ohms
|
|
Setup: Levels are measured across tip and ring with the line
open circuit. When performing the tests, connect the CPE to the
telephone line.
|
Setup: Levels are measured across tip and ring with the line
terminated by a 600 ohm resistor. When performing the tests,
replace the 600 ohm resistor by the CPE.
|
|
Example: Sending a FSK signal level of 120 mVrms requires a
CO voltage source level of 120 mVrms. The level across tip and
ring during testing is also 120 mVrms.
|
Example: Sending a FSK signal level of 120 mVrms requires a
CO voltage source level of 300 mVrms. The level across tip and
ring during testing will depend on the CPE’s off-hook
impedance. Since this is normally close to 600 ohms, the level
will be approximately 120 mVrms.
|
ETSI |
Document: ETS 300 778-1
Units: dBV
Source Impedance: not specified
Load Impedance: the CPE’s AC termination can be either:
a) greater than 8 kohms
b) impedance of 820 ohms in series with a parallel
combination of 360 ohms and 180 nF.
|
Document: ETS 300 778-2
Units: dBV
Source Impedance: not specified
Load Impedance: CPE’s AC termination requirements depend on
additional national standards
|
|
Setup: Levels are measured across tip and ring with the CPE
connected in the on-hook state (off-line).
|
Setup: Levels are measured across tip and ring with the CPE
connected in the off-hook state (on-line).
|
|
Example: Sending a FSK signal level of -20 dBV requires
adjusting the CO voltage source until the level across tip and
ring is -20 dBV. The actual CO sending level depends on the CPE’s
AC termination and the source impedance of the test setup.
|
Example: Sending a FSK signal level of -20 dBV requires
adjusting the CO voltage source until the level across tip and
ring is -20 dBV. The actual CO sending level depends on the CPE’s
AC termination and the source impedance of the test setup.
|